KARL FISCHER COULOMETRIC TITRATOR AZA1218

AZA1218 is a Karl Fischer coulometric titrator for trace moisture measurement in liquids, solids and gases. The AZA1218 features a 7-inch touchscreen, automatic drift compensation, adjustable stirring, data storage and a built-in thermal printer.

Description

Description

TECHNICAL SPECIFICATIONS

Parameter Specification
AZA LAB model AZA1218
Measurement method Karl Fischer coulometric method
Display measuring range 0.1 µg to 9999.9999 mg water
Water content range 0.000001%–100%; 0.001 ppm–10⁶ ppm (supplier-stated)
Resolution 0.1 µg water
Display 7-inch colour LCD touchscreen
Accuracy, 3–1000 µg water ±2 µg
Accuracy, above 1000 µg ±0.2% of measured value; excludes sampling error
Endpoint indication Screen, audible alert, printout and endpoint lamp
Blank / drift compensation Automatic microprocessor control
Stirring speed Adjustable
Delay titration 0–100 seconds
Electrolysis speed Peak 2.4 mg water/min
Electrolytic current 0–430 mA
Data and records Historical data storage; user-defined sample number
Printout Built-in thermal printer; 56 mm paper width
Self-check / clock Automatic fault diagnosis; persistent date/time
Optional function Automatic liquid filling and draining
Power supply 220 V ±10%, 50 Hz ±2.5 Hz; <100 W
Operating conditions 5–40 °C; relative humidity <85%
Instrument dimensions 350 × 260 × 198 mm
Net / gross weight 10 kg / 14.6 kg
Package dimensions 450 × 450 × 300 mm