AutoFlash – Automated Abel Flash Point Analyzer( AZA 1277)

AZA Lab’s AutoFlash-1277 is a fully automated, microcontroller-based Abel Flash Point Analyzer designed to accurately determine the flash points of combustible liquids. Conforming to IP 170, ISO 13736, and IS 1448 (P:20) standards, it ensures precise and safe analysis of samples in the temperature range of -30°C to 70°C, extendable up to 110°C.

Equipped with a built-in PC and touchscreen display, the AutoFlash-170 delivers intuitive operation, automated calibration, advanced ignition options, and rapid heating procedures for high-efficiency testing. Ideal for laboratories, petrochemical industries, and quality control applications, this standalone unit streamlines the testing process with enhanced accuracy, safety interlocks, and programmable procedures.

Description

Description

Key Features:

  • Standalone model with inbuilt PC and display
  • Accurate flash point determination as per global standards
  • User-friendly operation with programmable test cycles
  • Rapid heating to reach expected flash point efficiently
  • Auto search mode for unknown sample flash point detection
  • Interchangeable ignition source – gas or electric igniter
  • Automatic barometric pressure compensation
  • Built-in safety interlocks for reliable operation
  •  Step motor-controlled arm & stirrer for consistent results
Measurements:
Parameter Details
Auto Calibration Auto temperature calibration before every run – no external device required
Sample Temperature PT100 Class A, SS sheath, range: -100°C to 150°C
Bath Temperature PT100 Class A, disk type, range: -100°C to 200°C
Flash Detection Thermal Sensor
Ambient Pressure Built-in pressure sensor (700 to 1100 hPa / 500 to 800 mmHg)
System Configuration:
Function Specification
Heating/Cooling Cartridge heater (300W); Cryostat cooling to -30°C (optional)
Stirring Stepper motor controlled, 5–100 RPM user-defined
Ignition Source Electric/Gas igniter with 7-step brightness control, 0.5 to 5°C dip interval
Arm Movement Stepper motor swing arm with integrated cup cover, stirrer, sensors, igniter
Fire Detection Thermal sensor
Flame Extinguishing Auto flame cut-off on fire or flash detection
Safety Interlocks:
  • Temperature sensor (sample/bath) failure

  • Flash detector defect

  • Blown electric igniter

  • Arm not positioned correctly

  • Flash point exceeds selected or 400°C

  • Fire detection triggers alarm & auto shutdown