AUTOMATED CLEVELAND OPEN CUP FLASH POINT ANALYZER(AZA 1279)

AutoFlash-1278 is a programmable, fully automated, microcontroller-based standalone Cleveland Open Cup Flash Point Analyzer conforming to standards:
ASTM D92, ISO 2592, IS 1448 (P:69), IP 36.

It is suitable for all petroleum products, especially viscous materials, with flash points:

  • Above 79°C and

  • Below 400°C (excluding fuel oils)

Also used to determine the fire point, i.e., the temperature at which the test specimen sustains combustion for a minimum of 5 seconds.

Description

Description

Features:

  •  Standalone model with inbuilt PC & display
  •  Accurate flash point determination
  •  Simple operation: enter flash point, click start – fully automated process with result saving
  •  Rapid heating until temperature is just below expected flash point
  •  Inbuilt search mode for unknown samples
  • Interchangeable gas or electric igniter source

Methods & Interface:

  • Methods: ASTM D92, ISO 2592, IS 1448 (P:69), IP 3

  • Rapid heating with user-defined procedures

  • User Interface: Windows-based software with method storage

  • Display: 8.4” touchscreen, optional printer support

SYSTEM Configuration:
  • Heating: Cartridge Heater @ 300W
  • Cooling:
    • Peltier: +10°C to +110°C (Air-cooled to +25°C)
    • Refrigerated Circulator: -30°C to +110°C
  • Stirring: Stepper motor, 5–100 RPM
  • Ignition:
    • DC motor driven
    • Selectable electric/gas igniter
    • 0.5 to 5°C dipping intervals
    • Precise gas flow control
    • Auto gas igniter & 3-step electric igniter brightness
  • Arm Movement: DC motorized swing arm with fixed cup, sensors, and igniter
MEASUREMENTS:
  • Auto Calibration: Before each run, no external calibration device needed

  • Sample Temp: PT100 Class A, SS/Glass, -100°C to +150°C

  • Bath Temp: PT100 Class A, Disk type, -100°C to +200°C

  • Flash Detection: Thermal sensor

  • Fire Detection: UV sensor

  • Ambient Pressure: Built-in sensor, 700–1100 hPa

Reviews

There are no reviews yet.

Be the first to review “AUTOMATED CLEVELAND OPEN CUP FLASH POINT ANALYZER(AZA 1279)”