ASPHALT BENDING BEAM RHEOMETER (BBR) AZA1183

The AZA1183 Asphalt Bending Beam Rheometer (BBR) is designed to determine the low-temperature flexural creep stiffness of asphalt binder specimens under a controlled load in a temperature-regulated liquid bath.

The system measures specimen deflection and applied force to characterize binder stiffness and creep response. It is intended for asphalt binder evaluation and pavement-material research under the listed test methods.

 

Description

Description

KEY CAPABILITIES

  • Displacement sensor: 0–10 mm range, 1 µm resolution
  • Force sensor: 0–4900 mN range, 1 mN resolution
  • Temperature sensor: −50 °C to room temperature; ±0.1 °C accuracy
  • Liquid bath: −40 °C to room temperature; ±0.1 °C control accuracy
  • Creep stiffness measurement range: 20–1000 MPa
  • Single-specimen testing with 127 × 12.7 × 6.35 mm mold
  • USB data interface and 5 L stirred liquid bath

APPLICATIONS

Low-temperature creep stiffness evaluation of asphalt binders for pavement engineering, binder grading and research laboratories.

TECHNICAL SPECIFICATIONS

01 / MEASUREMENT & TEST SPECIFICATIONS

PARAMETER SPECIFICATION
Model AZA1183 — Asphalt Bending Beam Rheometer (BBR)
Test standards JTG E20 T0627 / AASHTO T313 / ASTM D6648 / EN 14771
Displacement sensor range 0–10 mm
Displacement sensor resolution 1 µm
Force sensor range 0–4900 mN
Force sensor resolution 1 mN
Temperature sensor range −50 °C to room temperature
Temperature sensor accuracy ±0.1 °C
Test mold size 127 × 12.7 × 6.35 mm
Specimen support top radius 3 ±0.3 mm
Loading axis contact Spherical; radius 6.3 ±1.3 mm
Number of specimens 1
Creep stiffness range 20–1000 MPa
Data interface USB
Liquid bath temperature −40 °C to room temperature
Liquid bath control accuracy ±0.1 °C
Liquid bath volume / stirring 5 L / 25 rpm
Main unit dimensions 610 × 610 × 620 mm
Host weight 80 kg
Host power supply AC 220 V, 1 kW
Power supply box AC 220 V, 1.8 kW
Refrigerator power supply AC 220 V, 3 kW

02 / LIQUID BATH & REFRIGERATION

  • Liquid bath operating range: −40 °C to room temperature
  • Temperature control accuracy: ±0.1 °C
  • Bath volume: 5 L
  • Stirring speed: 25 rpm.

03 / POWER & PHYSICAL DETAILS

  • Main unit: 610 × 610 × 620 mm; 80 kg
  • Host: AC 220 V, 1 kW
  • Power supply box: AC 220 V, 1.8 kW
  •  Refrigerator: AC 220 V, 3 kW.