BITUMEN PRESSURE AGING VESSEL (PAV) SYSTEM AZA1182

The AZA1182 Bitumen Pressure Aging Vessel (PAV) System is designed for laboratory conditioning of asphalt binders under controlled elevated temperature and air pressure to simulate long-term oxidative aging.

The system provides real-time temperature and pressure readings with graphical monitoring of their relationship during the aging process. The specified operating temperature range is 90–110 °C, with a nominal working pressure of 2.1 MPa ±0.1 MPa.

Description

Description

KEY CAPABILITIES

  • Working temperature range of 90–110 °C with ±0.5 °C control accuracy
  • Real-time temperature and pressure values with trend curves
  • Working pressure 2.1 MPa ±0.1 MPa
  • Actual pressure control accuracy ±0.02 MPa
  • Pressure sensor range 0–3 MPa; stated accuracy 0.5% (±0.015 MPa)
  • Automatic safety pressure relief above 2.5 MPa
  • Compressed-air cylinder connection with pressure-reducing valve and hose

APPLICATIONS

Used for pressure-aging conditioning of bituminous binders before subsequent laboratory characterization, following applicable test methods and conditioning procedures.

TECHNICAL SPECIFICATIONS

01 / OPERATING SPECIFICATIONS

PARAMETER SPECIFICATION
Model AZA1182 — Pressure Aging Vessel (PAV)
Test standards AASHTO R28 / ASTM D6521 / EN 14769
Working temperature range 90–110 °C
Temperature resolution 0.1 °C
Temperature control accuracy ±0.5 °C
Preheating time 40–60 min
Working pressure 2.1 MPa ±0.1 MPa
Actual pressure control accuracy ±0.02 MPa
Pressure sensor range 0–3 MPa
Pressure sensor accuracy 0.5% (±0.015 MPa)
Compressed-air supply Cylinder air pressure >3 MPa
Safety pressure relief Automatic above 2.5 MPa
Air connection Reducing valve and high-pressure conversion hose
Power supply 220 V, 50 Hz
Overall dimensions 550 × 350 × 450 mm
Monitoring Real-time temperature / pressure values and curves

02 / AIR SUPPLY & SAFETY

Requires compressed air from a cylinder at pressure greater than 3 MPa. Supplied with a high-pressure reducing valve and conversion hose for compatible industrial compressed-air cylinders. The safety valve automatically relieves vessel pressure when it exceeds 2.5 MPa.

03 / MONITORING & APPLICATION

Real-time display of temperature and pressure readings and their curve relationship • PAV conditioning of asphalt binders • Laboratory asphalt materials testing