ATOMIC ABSORPTION MERCURY ANALYZER AZA1344A

The AZA1344A Atomic Absorption Mercury Analyzer is a laboratory gas-phase mercury analysis instrument designed for sensitive and reliable measurement of mercury concentration. The system complies with the  Mercury Measuring Instrument Standard and incorporates a stable optical system, automated gas-flow control, and constant-temperature absorption cell technology.

The instrument maintains excellent baseline stability and provides reliable short-term and long-term measurement accuracy. Its constant-light optical path helps compensate for changes in the light source, improving long-term measurement stability.

A microcontroller automatically controls the solenoid valve to switch between the measurement state and cleaning state. During measurement, the system operates as a closed-loop gas path, while the cleaning process uses an open-loop gas path, facilitating efficient pipeline cleaning and stable measurement results.

The mercury gas continuously passes through the absorption cell, allowing measurement data to be obtained approximately every 2 seconds.

Description

Description

Key Features

  • Excellent baseline stability
  • High short-term and long-term measurement accuracy
  • Two-point calibration capability
  • Direct concentration reading
  • Suitable for gas, liquid, and solid samples
  • Constant-light optical path for improved long-term stability
  • Constant-temperature absorption cell for stable measurement
  • Automatic solenoid valve switching through microcontroller control
  • Automatic adjustment of cleaning gas flow
  • Fast measurement response
  • Does not require argon or nitrogen as carrier gas
  • Suitable for laboratories where argon or nitrogen supply is unavailable
  • Easy pipeline cleaning and maintenance

Technical Specifications

Parameter Specification
Measurement Principle Atomic Absorption Spectrometry
Measurement Range 0.1–100 µg/m³
Detection Limit 0.03 µg/m³
Precision Coefficient of Variation ≤ 5%
Linearity Correlation Coefficient r > 0.997
Relative Standard Deviation < 5%
Measurement Interval Approx. 2 seconds
Calibration 2-Point Calibration
Sample Types Gas, Liquid and Solid
Carrier Gas Requirement No Argon or Nitrogen Required
Optical System Constant-Light Optical Path
Absorption Cell Constant-Temperature Type
Gas Control Microcontroller-Controlled Solenoid Valve
Measurement Gas Path Closed-Loop
Cleaning Gas Path Open-Loop
Gas Flow Control Automatic Adjustment